Mitutoyo operates a policy of continuous improvement that aims to provide the customer with the benefit of the latest technological advances.
Therefore the company reserves the right to change any or all aspects of any product specification without notice.
D-4
D
About the charge function
The minimum illumination required in the
uncharged state is 60 lux.
As shown in the table at right, this SuperCaliper
can be used with confidence in a normal work
environment.
The charge function allows the operator to use
the SuperCaliper without interrupting work
even if the ambient illumination is temporarily
insufficient.
In the fully charged state this SuperCaliper
can operate for approximately one hour in an
environment of 50 lux illumination (less than the
minimum necessary illumination intensity).
The time necessary for full charge varies according
to the charging conditions. If the SuperCaliper
is left unused in an illumination of 500 lux
(usual for manufacturing environments), it takes
approximately one hour to reach full charge.
Air leak test equipment for water resistance inspection
Generally, an air leak test is used for evaluating water resistance.
Procedure: Place the measuring tool inside the capsule and seal it. Then fill the capsule and the test
chamber with air at the required pressure and close the isolating valves. If there is no leak in the
measuring tool, the differential pressure sensor will read zero, because the amount of air inside the test
chamber is unchanging. However, if there is a leak in the measuring tool, the differential pressure sensor
will show a non-zero reading due to a decrease in pressure inside the test chamber as air leaks into the
tool. By detecting this differential pressure, GO/NG judgement for the severity of the leak is performed.
This air leak test is performed for all coolant proof calipers and coolant proof micrometers.
Optional Accessories
(Only for models with data output function)
05CZA624: SPC data cable with pushbutton (1 m)
£58.50
05CZA625: SPC data cable with pushbutton (2 m)
£70.40
06ADV380A: USB Input Tool Direct USB-ITN-A (2 m)
£106.00
02AZD790A: SPC data cable for U-WAVE-T (160 mm)
£52.20
02AZE140A: SPC data cable for U-WAVE-T and footswitch
(160/500 mm)
£114.00
1500
1000
750
500
300
200
150
100
75
50
30
20
Site (possible operations)
Design room, drafting room
(Fine visual work)
Corridor, passage, stairs
(Very rough visual work)
Emergency staircase, warehouse
(Loading, unloading work)
Excerpts from JIS Z 9110 Artificial Illumination Intensity Standard
Conference room, control room
(Usual manufacturing environment)
(Normal visual work)
Machine room, electric room, lecture hall
(Rough visual work)
Air supply
Capsule
Isolating
valves
Isolating
valves
Test chamber
Differential
pressure sensor
Air leak test equipment for coolant proof caliper
Comentários a estes Manuais